CVE-2019-25719
Vulnerability data via NVD (ingested)
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
External references
Search for exposed instances
Shodan + Censys queries derived from NVD's CPE data. The vuln tag catches assets Shodan has explicitly linked to this CVE; the product / banner fingerprints find exposed instances even when the vuln tag was never applied (which is common).
More intel sources (5)
vuln:CVE-2019-25719vulnerabilities.cve_id: CVE-2019-25719CVE-2019-25719CVE-2019-25719"CVE-2019-25719" exploit -site:nvd.nist.gov